Subject: JEMX Internal Meeting 2004-04-13 Participants: ILR, CAO, JC, NL, NJW 1) Instrument status JEM-X2 was reactivated for 5 hours in rev. 179. Not enough for gain stabilization. Energyresolution @22 keV was about 11.3% so some has been regained (when switched off in rev. 170 it was 13.5%; nominal value is 9% e.g. in revolution 30) JEM-X1: Anode #83 has become unstable Electronic calibration alerts: Fast anode amplifier for JEM-X1 somewhat more noisy than in JEM-X2 Count rate alert limits in MOC had to be adjusted because on-board limits were increased. No secure data on development of energy resolution in JEM-X1 2) Laboratory tests postponed to next week. 3) Munich papers: 3 have been delivered 4) ISDC consortium: OSA4: COR, IMA, SPE improvements. Midisky, Spectral correction periods (IC files) /NJ ---------- Niels J. Westergaard Danish Space Research Institute Juliane Maries Vej 30, DK 2100 Copenhagen O Phone: +45 35 32 57 05; FAX: +45 35 36 24 75