Dear Carol Anne,

Here are my overheads for the minutes of the Stockholm SDAST meeting.

1) 041107_anod1.ps: diagram over JEM-X1 anodes on 041107 (red) compared with the same on 040911. this does not show any wild changes, but just a small widening of the new "death valley" around RAWX=185.

2) 041101_0_anod1.ps: Almost the same as previous but for a longer period (6 months) showing better the strong widening deepening of the "death valley", as well as the quite new dead node at RAWX=73, and the deepening of the old dead anode at RAWX=220.

3) Anodes_list.txt: Anodes status for both JEM-X instruments.

4) JMX1_Xe_time.ps: Shows the evolution of the measured position of the Xenon line from 1335 JEM-X1 detector's spectra (kindly supplied by PK) up to the revolution 245. (I did not shown this overhead in Stockholm but its equivalent up to rev. 229). A quicker rising of the Xenon line position is obviously observed since the beginning of July, probably related to a problem with one of the calibration sources.

5) JMX1_Xe_width.ps: Corresponding evolution of the detector resolution (absolute E in red, relative in green) for the Xenon line. The relative resolution rises up to 13% in average at the end of the period.

-- 
Dr. Jérôme CHENEVEZ                     Danmarks Rumcenter
Danish Space Research Institute         Juliane Maries Vej 30
http://www.dsri.dk/~jerome              DK-2100 Copenhagen Ø
Tel. +45 35 32 57 03                    FAX: +45 35 36 24 75