Gain Notes for Revolution 2668 Another revolution largely without data, very noisy and the whole is declared a BTI due to the inaccuracy of the energy determination and the extremely low gain and subsequent loss of events in the low-level discriminator. BTI = 8609.4 to 8612.0 IJD. This is one of the later revolutions in the mission where the calibration sources have become so weak that for the best energy determination an IC Gain History table is needed. These gain history tables are found automatically by OSA, or can be downloaded from the DTU FTP site. All offline gain/energy corrections (including the IC tables) are based on the two Fe calibration sources in JEM-X1, for both units, then corrected iteratively using the instrument background lines (Xe, backed up by Cu and Mo). Latest OSA software (version 10 and higher) automatically cuts out the first Science windows of data from each revolution to ensure that people do not use data from the instrument-settling period. This process can be overruled, but only by experienced users. JEM-X1 rating: very poor - hence BTI between 8609.4 to 8612.0 IJD. JEM-X2 rating: very poor - hence BTI. The JEM-X2 corrections are based on the JEM-X1 Fe calibration sources, since all the Cd sources are now too weak to use. CAO 02/08/2023