Gain Notes for Revolution 2664 The gain data and indeed the background line data is so weak for this revolution that it is not possible to perform gain corrections with any meaningful accuracy, though this has been tried. Therefore this revolution is declared a BTI for uncertain instrument parameters between 8599.0 and 8601.2 IJD This is one of the later revolutions in the mission where the calibration sources have become so weak that for the best energy determination an IC Gain History table is needed. These gain history tables are found automatically by OSA, or can be downloaded from the DTU FTP site. All offline gain/energy corrections (including the IC tables) are based on the two Fe calibration sources in JEM-X1, for both units, then corrected iteratively using the instrument background lines (Xe, backed up by Cu and Mo). Latest OSA software (version 10 and higher) automatically cuts out the first Science windows of data from each revolution to ensure that people do not use data from the instrument-settling period. This process can be overruled, but only by experienced users. JEM-X1 rating: Poor to indeterminate. JEM-X2 rating: Poor to indeterminate. The JEM-X2 corrections are based on the JEM-X1 Fe calibration sources, since all the Cd sources are now too weak to use. CAO 01/08/2023