Gain Notes for Revolution 1660 NB: This is an extremely short revolution that barely gets out of the settling period before a power cut on the satellite switches off all instruments. While there is very little data from this revolution, the gain correction for the available Science Windows is very good indeed. This is one of the later revolutions in the mission where the calibration sources have become so weak that for the best energy determination and IC Gain History table is needed. To get optimal results with this revolution it is necessary to use the IC gain history. See notes in revolutions prior to revolution 1433 to find out how to do this if you don't already know. All offline gain/energy corrections (including the IC tables) are based on the two Fe calibration sources in JEM-X1, for both units. Latest OSA software (version 10 and higher) automatically cuts out the first Science windows of data from each revolution to ensure that people do not use data from the instrument-settling period. This process can be overruled, but only by experienced users. JEM-X1: The Xe line analysis performed using the IC table shows excellent gain correction with all science windows (taken 1 at a time) having a Xe level within 1-2% of the ideal, except for the earliest settling points, and the last SCW which is completely misfitted due to lack of data. As usual, it is recommended that users avoid the first few science windows for energy-sensitive applications (automatically removed by OSA 10.0). JEM-X2: Identical comments as for JEM-X1. The JEM-X2 corrections are based on the JEM-X1 Fe calibration sources, since all the Cd sources are now to weak to use. CAO 7/04/2016