Gain Notes for Revolution 1016 This is one of the later revolutions in the mission where the calibration sources have become so weak that for the best energy determination and IC Gain History table is needed. JEM-X1: The Xe line analysis performed using the IC table shows good gain correction with all except the very first science windows having a Xe level within 2-3% of the ideal. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications. Calibration source number 4 (purple/orange) is no longer used for calibration purposes since an anode strip directly under the source has broken and no useable signal comes from this area of the detector any longer. JEM-X2: The Xe line analysis performed using the IC table shows excellent gain correction with all except the very first science windows having a Xe level within 1-2% of the ideal. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications. CAO 17/02/2011