Gain Notes for Revolution 960 JEM-X1: No scientific data or gain history for this unit. JEM-X2: This revolution is somewhat affected by the non-linear temperature effect on the gain of the instrument. The best energy correction is therefore achieved by using an IC table make offline at DNSI and delivered to ISDC which corrects for this effect. To get optimal results with this revolution it is necessary to use the IC gain history table that has been created offline at DNSC, instead of the gain history table created automatically by OSA. If you have the newest package of IC files from ISDC and these are correctly installed, OSA will find the IC gain history table automatically and you don't need to do anything. The updated ISDC Instrument Characteristics can be downloaded from http://www.isdc.unige.ch/integral/download/osa_sw However, if you don't have the IC gain table package from ISDC you can download the table from the gain history archive by anonymous ftp through the link provided. Download the table to some suitable subdirectory for your analysis, then set the hidden OSA parameter gainHist: gainHist="/". These files are used instead of the automatically generated gain history tables wherever there has been an unusual or non-linear behaviour of the detector gain. The Xe line analysis performed using the IC table shows excellent gain correction with all except the very first science windows having a Xe level within 1-2% of the ideal, with the vast majority of the science windows within 1% of the ideal. As usual however, it is recommended that users avoid the first few science windows for energy-sensitive applications. CAO 3/09/2010