Gain Notes for Revolution 799 JEM-X1: A quite normal revolution with fine gain fitting. All the data from this revolution should be useable. The Xe analysis confirms that the energy determination for this revolution is good, within about 2% of the ideal. However, as for all revolutions, the first few science windows should be avoided in any analysis requiring very good gain determination. Results shown here use OSA8.0 software and so include temperature variation in the smoothed gain model. The general low level of the Xe line is an artifact caused only by the high gain and high energy cutoff, and does not indicate an generally poor gain correction. Calibration source number 4 (purple/orange) is no longer used for calibration purposes since an anode strip directly under the source has broken and no useable signal comes from this area of the detector any longer. CAO 27/5/2009 JEM-X2: A quite normal revolution with fine gain fitting. All the data from this revolution should be useable. The Xe line analysis confirms the good gain calibration for this revolution with all except the very first Science Windows having a Xe line position with about 2% of the ideal value. However, it is suggested that users avoid the first few science windows in each revolution since this is where the instrument warms up and gain variations across the plate settle down. CAO 29/5/2009