Gain Notes for Revolution 707 JEM-X1: A quite normal revolution with fine gain fitting. All the data from this revolution should be useable. The Xe analysis confirms that the energy determination for this revolution is okay, within about 3% of the ideal. Since the temperature variation of the gain has tripled since the beginning of the mission, future offline calibrations, including temperature dependency, could possibly benefit this revolution when they become available in the coming year. JEM-X2: No scientific data or gain history data for this unit. CAO 5/08/2008