Gain Calibration Notes for Rev. 364 Relatively little noise and no large glitches. This revolution has one short break and one long break. The former has no effect on the gain, but the latter is sufficiently long as to produce a steep gain increase when the High Voltage is switched on again, just as we see at the beginning of each revolution, after perigee. Interestingly, the most noisy, glitch source, number 4 is most strongly affected by this effect. The smoothing is however correct, so there is no problem using data from this entire revolution. The Xenon line analysis shows that the Xe line is well within the 29-30keV range for most of the revolution. The only time period where the Xe line was outside the 29-30 keV range was at the very beginning of the revolution just after switch on. Users should be cautious when using these first science windows in a revolution for critical spectral analyses. Otherwise, the gain correction has worked very well. CAO 7/2/2007