Gain Notes for Revolution 255 JEM-X1: A quite normal revolution with fine gain smoothing. All data from this revolution should be useable. The first two points of each data set are the partial switch-on with the High Voltage 3 steps below its nominal value. Nice clean glitch on source 1. Cadmium sources clearly show competing fast and slow relaxation effects that produce a little hook at the beginning of the revolution. Xe Analysis: the evolution of the Xe line position indicates that there is no hidden gain suppression on the plate despite a clear burst of solar activity in the middle of the revolution. There is however, clear evidence of an uncorrected temperature signal in the data, and this will be dealt with in the next version of OSA (i.e. the version after 6.0). CAO 17/11/2006