Gain Notes for Revolution 37 JEM-X1: No scientific data or gain history for this unit. JEM-X2: A quite normal revolution with fine gain fitting. All the data from this revolution should be useable. However, all the science data in this revolution is in Restricted Imaging mode, which is only partially supported by the OSA software, and no Xe analyis has been done of this data. A full Xe line analysis is required to confirm that the gain calibration has behaved correctly. First few points on each curve belong to the partial switch-on for which HV is set 3 steps below its nominal value. As always, however, it is suggested that users avoid the first few science windows in each revolution since this is where the instrument warms up and gain variations across the plate settle down. CAO 11/12/2007